Picture of Anesthesia Machine

Anesthesia Machine

• Precise and reliable delivery of oxygen and anesthetic gases • Comprehensive monitoring, control, and mechanical support of patient respiration (ventilation) • Enhanced patient safety ensured through integrated monitoring and alarm systems • Designed for use in a broad range of general and specialized surgical procedures

$0.000
Sharing

Misc

Operating principle Bag-in-bottle
Pressure Control (PCV) 2 – 100 cm H2O
Pressure Support (PSV) 0 – 100 cm H2O
Limited Pressure in TCPL 3 – 70 cm H2O
Resulting Minute Volume 0.01 – 17 L/min
Respiratory Rate 1 – 150 bpm
Continuous Flow (NEO-INF) 2 – 40 L/min
Limited Pressure in TCPL (NEO-INF) 3 – 70 cm H2O
Maximum Pressure Limited (Safety Limits) Up to 120 cm H2O
Inspiratory Time 0.5 – 3 s (30 s in APRV)
PEEP/CPAP 0 to 50 cm H2O
Inspiratory Flow 0.2 – 180 (L/min)
I:E Ratio (Volume Mode) 1:1
, 1:20
I:E Ratio (Pressure Control) 5:1 , 1:20
Expiratory Sensitivity 5% to 80% of Peak Flow
Inspiratory Sensitivity (Below PEEP) Flow Triggered (0.2 – 20 L/min)
, Pressure Triggered (0.5 – 20 cm H2O)
Pause Insp/Exp (VCV Only) 0 to 2 Seconds With 0.25 s Increments
Other Properties Nebulizer , Auto PEEP (total PEEP / auto PEEP)
, Static and Dynamic Compliance
, Inspiratory and Expiratory Resistance
, Imposed Work of Breathing (WOBi)
, Trapped Volume (Tidal Volume/Total Volume/Trapped)
, Rapid Shallow Breathing Index (F/VT Index)
, Air Way Pressure (Peak, Platea, Mean, PEEP)
, I:E Ratio Current
, Expired Time Constant (TCexp)
, Expired Tidal Volume
, O2 Percentage
, Inspiratory Time (Ti) and Expiratory Time (Te)
, Rate (Total rate Expressed in Breaths per Minute)
, Expired Minute Volume
, Volume Correction according to patient Circuit Compliance
, Leak Compensation available in all operative's modes
, Body Temperature Volume Correction (BTPS)
, Altitude Compensation for Volume Correction
, Tidal Volume Setting Based on Ideal Body Weight (IBW)